IB-11E Incubator (150L)
Structural Functional Features
- Economical model series that minimizes the purchase burden.
- An incubator in which the internal chamber is heated using a film heater on the outer wall of the space where the sample is placed.
- The sample space is independently kept clean and the structure protects against sources of contamination.
- External door can be opened and closed smoothly with magnetic packing to minimize sample impact.
- Stainless steel interior and shelves are excellent in terms of corrosion resistance and clean maintenance.
Use Convenience Features
- Accurately controlling temperature based on temperature calibration / auto-tuning / microprocessor PID.
- The inner glass door is completely enclosed with silicone packing, making it easy to observe the sample without changing the temperature.
- Save and use 3 frequently used temperatures.
- Wait On/Off timer. (up to 99 hours 59 minutes)
- Provides notification/alarms when the door is open for an extended period of time.
Outstanding Safety
- Top-rated overheating protection system.
(registration 10- 0397583) - Over temperature limit function.
- Malfunction prevented by controller lock function.
Chamber volume (L / cu ft) | 150 / 5.3 |
---|---|
Temperature | |
Range (℃ / ℉) | Amb. +3 to 60 / Amb. +5.4 to 140 |
Fluctuation at 37℃ (±℃ / ℉) | 0.3 / 0.54 |
Variation at 37℃ (±℃ / ℉) | 0.8 / 1.44 |
Heating time to 37℃ (min.) | 22 |
Recovery time at 37℃ (min.) | 2 |
Dimension | |
Interior (W x D x H, mm / inch) | 500 x 515 x 585 / 19.7 x 20.3 x 23 |
Exterior (W x D x H, mm / inch) | 595 x 630 x 865 / 23.5 x 24.8 x 34 |
Net weight (Kg / lbs) | 47 / 103.6 |
Shelves | |
Quantity of shelves (standard/max.) | 2 / 13 |
Max. Load per shelf (Kg / lbs) | 30 / 66 |
Electrical requirements (120V, 60Hz, A) | 4.2 |
Cat. No. | AAH26026U |
※ Technical data (according to DIN 12880 before 2013)
Voltage | Hertz | Phase | Plug Type |
120 | 50/60 | 1P | Type B l NEMA 5 |
Reviews