OF4-05PW Programmable Forced Convection Oven (Window, 61L)
Easy setting program temperature for everyone
- User interface that provides a quick view of temperature changes
(heating, stable, and cooling) and fan speed also. - Set various test condition using 10 programs and 10 steps, which can be repeated up to 99 times.
- Start temperature, temperature and time for each step, and Opti-flowTM all adjustable as desired.
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img1-300x216.jpeg)
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img2-300x216.jpeg)
Structural and functional features for improved user convenience
- Opti-flow™, a function to flexibly adjust the fan speed
according to the specimen type and test conditions. - Stacking Kit (optional) ensuring the safe fitting of one unit
to another, thus doubling the amount of work done in a given area for improved space efficiency. - Hands-free™ door system to allow users to easily open
the door while carrying samples with both hands.
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img3-300x216.jpeg)
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img4-300x216.jpeg)
Real-time monitoring to check status of device
- BMS communication port that helps monitor the device temperature on the building management system, enabling quick response in case of abnormality. (option)
- Monitor and control via mobile app anytime, anywhere with LC Connected. (mobile monitoring system) (option)
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img5-300x216.jpeg)
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img6-300x216.jpeg)
Convenient checking for the test progress and data
- RS-232/USB port and compatible software provided to enable easy device control and data management using PC.
- Test data is automatically saved and easily downloaded via USB;
easy to edit and use. (in CSV format) - 36 latest events, such as errors and door opened or closed, automatically recorded, helping user obtain feedback on experimental results.
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img7-300x216.jpeg)
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img8-300x216.jpeg)
Reliable temperature control performance based on stable control mechanisms
- Wide temperature range of up to 250℃.
- Three-point temperature calibration that ensures high temperature accuracy over a wide temperature range.
- Ramping rates (℃/min) flexibly adjustable for each test performed.
- Excellent temperature distribution through the formation of optimal airflow for uniform heat transfer.
(±1.9℃ or below at 100℃)
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img9-300x216.jpeg)
![](https://labcompanionshop.com/wp-content/uploads/2022/12/OF4-P_img10-300x216.jpeg)
Model | Solid door (without window) | OF4-05P | ||||||
---|---|---|---|---|---|---|---|---|
Window door | OF4-05PW | |||||||
Chamber volume (L/cu ft) | 61/2.15 | |||||||
Temperature 1) | Range (℃/℉) | Amb. + 10~250 / Amb. + 18 ~ 482 | ||||||
Fluctuation at 100℃ (±℃/±℉) | 0.3/0.54 | |||||||
Variation at 100℃ (±℃/±℉) | 1.9/3.42 | |||||||
Heating time to 100℃ (min) | 10 | |||||||
Recovery time at 100℃ (min) | 4 | |||||||
Control Panel | 5inch TFT Color LCD | |||||||
Communication interface | USB, RS-232, BMS | |||||||
Function | Program setting | 10 program, 10 step, start temp. | ||||||
Scehdular | Set day, time, and program No | |||||||
Recently event record | 36ea | |||||||
Fan speed adjustable | Three-step wind speed settings | |||||||
Recommended temperature | Three frequently used temperature settings suggested | |||||||
Graph | Available checking the stored and real-time data as graphs | |||||||
Data storage | Save the test data in CSV format | |||||||
Temp. ramping rate (℃/min. / ℉/min.) | 0~9/0~16.2 | |||||||
Dimensions | Interior (W x D x H, mm/inch) | 380 x 330 x 480 / 14.97 x 13.0 0x 18.90 | ||||||
Exterior (W x D x H, mm/inch) | 558 x 648 x 684 / 21.97 X 25.51 X 26.93 | |||||||
Net weight (kg/lbs) | 53/116.8 | |||||||
Shelves | Quantity of shelves (Standard/max.) | 2/5 | ||||||
Max. Load per shelf (kg/lbs) | 30/66.1 | |||||||
Number of Air change per hour at 100 ℃ 2) | 22 | |||||||
Electrical requirements (120V, 60Hz, A) | 12.3 | |||||||
Cat. No. | Solid door (without window) | AAH12926U | ||||||
Window door | AAH12966U |